Triamterene Compound

 

 

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: K2EDTA

Storage Temperature: -20 °C

LLOQ: 2 ng/mL

ULOQ: 384 ng/mL

Sample Volume: 100 mcL

Precision and Accuracy:

  • LLOQ QC: 4.8% CV; -6.5% Bias
  • Low QC: 3.2% CV; 2.3% Bias
  • Mid QC: 1.5% CV; 0.3% Bias
  • High QC: 2.0% CV; -3.9% Bias
  • Above Range QC (5x Dilution): 1.5% CV; -7.9% Bias

Stability:

  • Short-Term (Bench-Top): 25 hrs 40 min
  • Long-Term: 124 days at -20 °C
  • Freeze/Thaw Cycles: 3

Comments: 

Method Number: V2291P1