Method: Ion-Selective Electrode

Matrix: Human Urine

Storage Temperature: -20 °C

LLOQ: 1 mM

ULOQ: 120 mM

Sample Volume: 500 mcL

Precision & Accuracy:

  • LLOQ QC: 4.9% CV; -11.5% Bias
  • Low QC: 5.2% CV; -2.0% Bias
  • Mid QC: 3.8% CV; -3.8% Bias
  • High QC: 4.3% CV; -1.9% Bias
  • Above Range QC (2x Dilution): 3.4% CV; 4.0% Bias

Stability:

  • Short-Term (Bench-Top): 24 hours
  • Long-Term: 48 days at -20 °C
  • Freeze/Thaw Cycles: 3

Method Number: V2340U1