Noroxycodone in Human Plasma Assay

 

 

 

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: Sodium Heparin

Storage Temperature: -20 °C

LLOQ: 0.25 ng/mL

ULOQ: 50 ng/mL

Sample Volume: 500 mcL

Precision & Accuracy:

  • LLOQ QC: 13.0% CV; -8.0% Bias
  • Low QC: 6.4% CV; -6.0% Bias
  • Mid QC: 3.3% CV; -3.4% Bias
  • High QC: 5.1% CV; -2.9% Bias

Stability:

  • Short-Term (Bench-Top): 4 hours and 15 minutes
  • Long-Term: 20 weeks at -20 °C
  • Freeze/Thaw Cycles: 3

Comments: Assay quantitates Oxycodone and Oxymorphone (0.25 – 50 ng/mL each)

Method Number: V2043P1