Morphine 6G in Human Plasma Assay

 

 

 

 

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: Potassium EDTA

Storage Temperature: -70 °C

LLOQ: 2 ng/mL

ULOQ: 384 ng/mL

Sample Volume: 250 mcL

Precision & Accuracy:

  • LLOQ QC: 4.8% CV; 1.0% Bias
  • Low QC: 3.5% CV; 0.4% Bias
  • Mid QC: 3.1% CV; 0.5% Bias
  • High QC: 2.2% CV; -0.6% Bias
  • Above Range QC (5x Dilution): 3.7% CV; 1.0% Bias

Stability:

  • Short-Term (Bench-Top): 25 hours and 15 minutes
  • Long-Term: 622 days at -70 °C
  • Freeze/Thaw Cycles: 4

Method Number: V2323P1