Molindone in Human Plasma Assay

 

 

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: Sodium Heparin

Storage Temperature: -20 °C

LLOQ: 0.250 ng/mL

ULOQ: 100 ng/mL

Sample Volume: 200 mcL

Precision & Accuracy:

  • LLOQ QC: 8.5% CV; -0.1% Bias
  • Low QC: 5.1% CV; 2.5% Bias
  • Mid QC: 3.5% CV; -0.5% Bias
  • High QC: 4.0% CV; -6.2% Bias
  • Above Range QC (5x Dilution): 5.9% CV; 11.3% Bias

Stability:

  • Short-Term (Bench-Top): 4 hours and 40 minutes
  • Long-Term: 103 days at -20 °C
  • Freeze/Thaw Cycles: 4

Method Number: V2271P1