Method: ELISA

Matrix: Human Plasma

Anticoagulant: K2EDTA

Storage Temperature: -80 °C

LLOQ: 30 pg/mL

ULOQ: 2000 pg/mL

Sample Volume: 200 mcL

Precision & Accuracy:

  • LLOQ QC: 7.1% CV ; -9.4% Bias
  • Low QC: 2.1% CV ; -8.3% Bias
  • Mid QC: 2.3% CV ; 5.3% Bias
  • High QC: 1.7% CV ; -1.0% Bias
  • Above Range QC (5x Dilution): 6.3% CV ; -13.5% Bias

Stability:

  • Short-Term (Bench-Top): 18 hours
  • Long-Term: 320 days at -80 °C
  • Freeze/Thaw Cycles: 3 cycles

Method Number: Q2392P1