Haloperidol in Human Plasma Assay

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: Potassium EDTA

Storage Temperature: -20 °C

LLOQ: 25 pg/mL

ULOQ: 6400 pg/mL

Sample Volume: 300 mcL

Precision & Accuracy:

  • LLOQ QC: 6.5% CV; -0.1% Bias
  • Low QC: 2.9% CV; 2.2% Bias
  • Mid QC: 2.0% CV; -0.3% Bias
  • High QC: 1.6% CV; 0.1% Bias
  • Above Range QC (5x Dilution): 1.2% CV; -1.2% Bias

Stability:

  • Short-Term (Bench-Top): 21 hours and 30 minutes
  • Long-Term: 120 days at -20 °C
  • Freeze/Thaw Cycles: 4

Method Number: V2331P1