Flurbiprofen in Human Urine Assay

 

 

 

 

Method: HPLC/Fl

Matrix: Human Urine

Storage Temperature: -20 °C

LLOQ: 0.5 mcg/mL

ULOQ: 20 mcg/mL

Sample Volume: 50 mcL

Precision & Accuracy:

  • LLOQ QC: R(-) 14.8% CV; 8.0% Bias; S(+) 11.3% CV; 6.0% Bias
  • Low QC: R(-) 14.7% CV; 7.1% Bias; S(+) 12.3% CV; 4.3% Bias
  • Mid QC: R(-) 6.6% CV; 4.7% Bias; S(+) 5.0% CV; 1.2% Bias
  • High QC: R(-) 8.6% CV; 3.2% Bias; S(+) 7.6% CV; -0.1% Bias

Stability:

  • Long-Term: 15 months at -20 °C
  • Freeze/Thaw Cycles: 5

Method Number: V0891U1