Fenofibric Acid in Human Plasma Assay

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: Sodium Heparin

Storage Temperature: -20 °C

LLOQ: 10 ng/mL

ULOQ: 5,000 ng/mL

Sample Volume: 100 mcL

Precision & Accuracy:

  • LLOQ QC: 4.8% CV; -6.7% Bias
  • Low QC: 3.1% CV; 3.6% Bias
  • Mid QC: 3.5% CV; 0.5% Bias
  • High QC: 3.4% CV; 0.1% Bias
  • Dilution QC (High QC, 4x): 7.8% CV; 6.3% Bias

Stability:

  • Short-Term (Bench Top): 24 hours
  • Long-Term: 149 days at -20 °C
  • Freeze/Thaw Cycles: 3

Comments: Stable in whole blood for at least 1 hour at room temperature

Method Number: V2350P1