Capture

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: K2EDTA

Storage Temperature: -20 °C

LLOQ: 0.25 ng/mL

ULOQ: 50 ng/mL

Sample Volume: 500 mcL

Precision & Accuracy:

  • LLOQ QC: 12.5% CV; -4.0% Bias
  • Low QC: 5.3% CV; 0.0% Bias
  • Mid QC: 5.0% CV; -1.4% Bias
  • High QC: 6.7% CV; -0.6% Bias
  • Above Range QC (10x Dilution): 3.4% CV; -1.7% Bias

Stability:

  • Short-Term (Bench-Top): 3.58 hours
  • Long-Term: Not established
  • Freeze/Thaw Cycles: 2

Comments: Also validated for Enalprilat

Method Number: V2048P1