Benazeprilat Human Plasma Assay Performance

 

 

 

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: Sodium Heparin

Storage Temperature: -20 °C

LLOQ: 2 ng/mL

ULOQ: 500 ng/mL

Sample Volume: 200 mcL

Precision & Accuracy:

  • LLOQ QC: 7.3% CV; 4.4% Bias
  • Low QC: 4.8% CV; -3.6% Bias
  • Mid QC: 4.8% CV; -0.2% Bias
  • High QC: 5.7% CV; 0.7% Bias
  •  Above Range QC (5x Dilution): 4.2% CV; -4.1% Bias

Stability:

  • Short-Term (Bench-Top): 4 hours
  • Long-Term: 11 weeks at -20 °C
  • Freeze/Thaw Cycles: 3

Method Number: V2077P4