Benazepril-in-human-plasma

 

Method: LC-MS/MS

Matrix: Human Plasma

Anticoagulant: Sodium Heparin

Storage Temperature: -20 °C

LLOQ: 2 ng/mL

ULOQ: 500 ng/mL

Sample Volume: 200 mcL

Precision & Accuracy:

  • LLOQ QC: 11.1% CV; 0.6% Bias
  • Low QC: 7.0% CV; 1.9% Bias
  • Mid QC: 7.4% CV; 1.8% Bias
  • High QC: 6.7% CV; -4.5% Bias
  • Above Range QC (5x Dilution): 2.7% CV; 2.6% Bias

Stability:

  • Short-Term (Bench-Top): 4 hours
  • Long-Term: 11 weeks at -20 °C
  • Freeze/Thaw Cycles: 3

Method Number: V2077P4